RELATIVE OUTPUT FACTOR (ROF) OF GAMMA KNIFE’S MEASUREMENT USING IONIZATION CHAMBER AND GAFCHROMIC FILMS

nazaroh nazaroh, Okky Agassy Firmansyah, Ade Rizky Setiadi, Irhas Irhas, Elia Sudiatmoko

Abstract


The important things on Leksell Gamma Knife (LGK) is determination of Relative Output Factor (ROF), which is required for beam data collection (BDC) process, as an input value to the Treatment Planning System (TPS). The ROF is also needed to traceability of dosimetry measurements and patient safety. ROF measurements on LGK using PTW PinPoint 3D ionization chamber: 0.016 cm3, PTW Semiflex: 0.125 cm3, and Exradin A16: 0.017 cm3 at collimators 4, 8 and 16 mm. and gafchromic films. ROF values obtained in hospital-A, using PTW PinPoint 3D was: [(1.00 ± 0.002)], [(0.80 ± 0.02]); [(0.42 ± 0.002)]; and using Exradin A16 [(1.00 ± 0]; [(0, 87 ± 0)]; [(0.69 ± 0.0002)]. And using gafchromic films was: [(1.00 ± 0.006]; (0.96 ± 0.005); (0.86 ± 0.06)]; respectively for collimators 4, 8 and 16 mm, and Measurement of ROF in hospital-B using using a PTW PinPoint 3D was: [(1 ± 0); (0.75 ± 0); and (0.33 ± 0)] and using Exradin A16 was: [(1; 0.87; and 0.69)] and gafchromic films was: [(1.00 ± 0.03); (0.98 ± 0.045); and (0.85 ± 0.09)]; The results of ROF’s measurement were compared with the ROF values of Novotny, et al (2009) and TPS. The Gafchromic’s ROF values were relatively closer than both of them, while the ROF value of ionization. chamber at 4 mm collimators was relatively lower, because measurement in a small field occurs an averaging effect volume and lack of lateral charge particle equilibrium (LCPE).


Keywords


relative output factor; leksell gamma knife; ionization chamber; small field; gafchromic film

Full Text:

PDF

References


Araki, Fujio, Tohru Ikegami, Tatsuyo Ishidoya, and Hideo D. Kubo. (2003), “Measurements of Gamma-Knife Helmet Output Factors Using a Radiophotoluminescent G.ass Rod Dosimeter and a Diode Detector.” Medical Physics 30(8):1976–81.

Baffa, Oswaldo, and Angela Kinoshita. (2014), “Clinical Applications of Alanine/Electron Spin Resonance Dosimetry.” Radiation and Environmental Biophysics 53(2):233–40.

Bhatnagar, Jagdish P., Josef Novotny, Ajay Niranjan, Douglas Kondziolka, John Flickinger, Dade Lunsford, and M. Saiful Huq. (2009), “First Year Experience with Newly Developed Leksell Gamma Knife ® Perfexion TM Full Text Introduction.” 34–36.

Bouchard, Hugo, Jan Seuntjens, Simon Duane, Québec Hl, and Hugo Palmans, (2015), “Detector Dose Response in Megavoltage Small Photon Beams . I . Theoretical Concepts.” 42(10):6033–47.

Das, Indra J., George X. Ding, and Anders Ahnesjö, (2008). “Small Fields: Nonequilibrium Radiation Dosimetry.” Medical Physics 35(1):206–15.

Firmansyah, Assef Firnando, C. Tuti Budiantari, Gatot Wurdiyanto, and Nurman Rajagukuguk. (2016), “Stabilitas Luaran Berkas Radiasi Cobalt 60 Pesawat Pisau Gamma Leksell Perfexion.” Pp. 193–99 in Seminar Nasional XII SDM Teknologi Nuklir.

Firmansyah, Assef Firnando, Okky Agassy Firmansyah, Nurman Rajagukuguk, and Dea Ryangga. (2020), “Studi Awal Pengukuran Daisy Chaining Detektor Ionisasi Pada Berkas Foton 6 MV Flattening Filter Free Lapangan Kecil.” Jurnal Ilmu Fisika (JIF) 12(1):16–25.

Firmansyah, Assef Firnando, Sri Inang Sunaryati, Nurman Rajagukuguk, and Gatot Wurdiyanto. (2017), “Penentua Dosis Serap Air Berkas Radiasi Co-60 Pesawat Pisau Gamma Leksell Perfexion No. Seri 6428.” Pp. 63–66 in Seminar Keselamatan Nuklir.

Gonzalez-Lopez, Antonio, Juan Antonio Vera-Sanchez, and Jose Domingo Lago-Martin,. (2015), “Small Fields Measurements with Radiochromic Films.” Journal of Medical Physics 40(2):61–67.

Hassani, Hossein, Hassan Ali Nedaie, Mohammad Hassan Zahmatkesh, and Kaveh Shirani, (2014), “A Dosimetric Study of Small Photon Fields Using Polymer Gel and Gafchromic EBT Films.” Medical Dosimetry 39(1):102–7.

Higuchi, Yoshinori, Shinji Matsuda, and Toru Serizawa, (2017), “Gamma Knife Radiosurgery in Movement Disorders: Indications and Limitations.” Movement Disorders 32(1):28–35.

Hrsak, Hrvoje, Marija Majer, Timor Grego, Juraj Bibic, and Zdravko Heinric, (2014), “Correction of Measured Gamma-Knife Output Factors for Angular Dependence of Diode Detectors and PinPoint Ionization Chamber Hrvoje Hr.” Physica Medica 30(8):914–19.

Klawikowski, Slade J., James N. Yang, John Adamovics, and Geoffrey S. Ibbott, (2014). “PRESAGE 3D Dosimetry Accurately Measures Gamma Knife Output Factors.” Phys. Med. Biol 59:N211–20.

Li, Yinghui, Lixin Chen, Jinhan Zhu, and Xiaowei Liu, (2017), “The Combination of the Error Correction Methods of Gafchromic EBT3 Film.” PLoS ONE 12(7):1–17.

Ma, Lijun, Jonas Johansson, Lynn Verhey, Per Kjall, Josef Novotny Jr, and Hakam Nordstrom, (2009), “A Simple and Effective Method for Validation and Measurement of Collimator Output Factors for Leksell Gamma Knife ® Perfexion TM.” Phys. Med. Biol 54(12):3897–3907.

Morales, Johnny E., Martin Butson, Scott B. Crowe, Robin Hill, and J. V. Trapp, (2016), "An Experimental Extrapolation Technique Using the Gafchromic EBT3 Film for Relative Output Factor Measurements in Small X-Ray Fields.” Medical Physics 43(8):4687–92.

Mourya, Ankur, Lalit Mohan Aggarwal, Abhijit Mandal, Uday Pratap Shahi, and Chhape Ram, (2018), “An Inexpensive Method of Small Photon Field Dosimetry with EBT3 Radiochromic Film.” Health and Technology 8(6).

Nazaroh, Assef Firnando Firmansyah, and Nurman Rajagukguk, (2012), “Leksell Gamma Knife (LGK) Perfexion/6121, Pertama Di Indonesia Dan Verifikasi Keluarannya.” pada Pertemuan dan Presentasi Ilmiah Standardisasi (PPIS).

Novotny, Josef, Jagdish P. Bhatnagar, Jonas Johansson, and Greg Bednarz. (2010), “Assessment of Variation in Elekta ® Plastic Spherical-Calibration Phantom and Its Impact on the Leksell Gamma Knife ® Calibration.” Medical Physics 37(9):5066–71.

Novotny, Josef, Jagdish P. Bhatnagar, Mubina A. Quader, Greg Bednarz, L. Dade Lunsford, and M. Saiful Huq, (2009), “Measurement of Relative Output Factors for the 8 and 4 Mm Collimators of Leksell Gamma Knife Perfexion by Film Dosimetry.” Medical Physics 36(5):1768–74.

Ulya, Syarifatul, Wahyu Edy Wibowo, Nuruddin Nuruddin, and Supriyanto Ardjo Pawiro. (2016), Karakteristik Dosimetri Film Gafchromic Ebt3 Pada Berkas Elektron Dan Foton.” Spektra: Jurnal Fisika Dan Aplikasinya 1(2):149–56.

Zeverino, Michele, Maud Jaccard, and David Patin, (2017), Commissioning of the Leksell Gamma Knife ® Icon TM.” Medical Physics 44(February):355–63.